The high resolution of the diffractometer, together with its ability to measure intensities accurately, allows the whole profiles of the diffraction lines to be studied. The width of the profiles is dependent on the spectral width of the characteristic radiation and on the geometry of the instrument. It is shown theoretically that the true Bragg angles can only be found, for work of the highest accuracy, from the centroids (first moments) of the profiles, the mathematics required to correct the peak positions being prohibitive. Experimental procedures required to put this theoretical advantage of the centroid into practice are explained, and difficulties due to the slow decay of the tails of the lines, and the uncertainty of background level are circumvented. An expression for the uncertainty in the centroid, found by step-by-step scanning by the fixed-time technique, due to counting statistics, is derived. This is plotted for the case of a single Cauchy profile. It is found that this uncertainty depends greatly on the peak-to-background ratio, and hence that scintillation or proportional counters have a marked advantage over Geiger counters for this type of work. It is concluded that, when the maximum accuracy in lattice-spacing measurements is desired, the centroid must be used instead of the peak, and that this does not involve excessive inconvenience in practice.