Key result
An 8-minute cardiac 31P-MRS protocol yields equivalent PCr/ATP ratios to standard ≥20-minute protocols.
Why the study?
Long acquisition times of 20 minutes or more required for reliable cardiac 31P-magnetic resonance spectroscopy have limited its clinical utility in patients with severe cardiac disease.
Does a short 8-minute 31P-MRS protocol provide equivalent PCr/ATP ratios compared to a standard 20-minute protocol in healthy volunteers?
Observational (n=22)
No
Does a short 8-minute 31P-MRS protocol provide equivalent PCr/ATP ratios compared to a standard 20-minute protocol in healthy volunteers?
A novel 8-minute cardiac 31P-MRS protocol provides equivalent PCr/ATP ratios to the standard 20-minute protocol, facilitating its use in patients intolerant of long scan times.
Answers are grounded in this paper and cite their sources.
Supports abbreviated protocols in volunteers; leaves open clinical validation in patients.
Introduction Cardiac 31P-MRS is the only non-invasive in vivo technique for the determination of cardiac high energy phosphate metabolism. Changes in cardiac phosphocreatine to adenosine triphosphate ratios (PCr/ATP) occur in common cardiac pathologies and have diagnostic, prognostic and therapeutic utility. However, long acquisition times (20 minutes or more, depending on heart rate) required to achieve sufficient signal to noise ratios for reliable interpretation have limited the clinical utility of 31P-MRS studies in patients with severe cardiac disease. We have developed an 8 minute 31P-MRS protocol and demonstrate the validity of this 'short' acquisition by comparison with a 'long' (at least 20 minutes) method of published reproducibility (Tyler, NMR Biomed:2008). Purpose To design a robust, 'short' cardiac 31P-magnetic resonance spectroscopy (31P- MRS) protocol which facilitates acquisition within a clinically acceptable timeframe. Methods Protocol development This 'short' protocol essentially incorporates a larger voxel (93 mls compared to 39 mls for 'long' protocol) but eliminates extra myocardial contamination by: active suppression of chest wall muscle and liver signals, raw data acquisition weighted to reduce contamination arising from outside the nominal voxel. The accuracy of the data and its interpretation is improved by: optimised radio frequency (RF) pulse, flip angle calibration (at voxel of interest) used during post-processing to calculate and correct for subject variation to coil loading. calibrated signal enhancement (Nuclear Overhauser Enhancement (NOE)), rapid repetition time (with calibrated saturation correction). Validity 22 healthy volunteers (age 42 ± 16.5; 13 males, 9 females), were scanned (3 T Siemens Trio) with both the 'long' and 'short' acquisitions. Acquisition parameters for both protocols are summarised and compared in Table 1. Results There was no difference of derived PCr/ATP ratios for both methods ('short' 1.83 ± 0.32; 'long' 1.78 ± 0.27). Bland-Altman analysis demonstrates excellent agreement between the two methods (Figure 1) confirming equivalence for clinical purposes. Figure 2 shows an example of spectra acquired from one subject, using both methods. Conclusion We have developed a novel 'short' cardiac 31P-MRS protocol of high data quality. This protocol allows cardiac spectroscopy to be measured in patients who are often intolerant of long acquisition times, such as those with severe cardiac disease and children. Hence this work provides a useful tool for the routine clinical assessment of cardiac 31P-MRS.
No takes yet. Share an insight, caveat, or question.
Dass et al. (2010) conducted an observational in Healthy volunteers (n=22). Short (8-minute) cardiac 31P-MRS protocol vs. Long (≥20-minute) cardiac 31P-MRS protocol was evaluated on Derived cardiac phosphocreatine to adenosine triphosphate (PCr/ATP) ratio. An 8-minute cardiac 31P-MRS protocol yielded PCr/ATP ratios equivalent to a standard ≥20-minute protocol in healthy volunteers.