Demonstrates that pacemaker-mediated tachycardia can be triggered by abnormal sensing of ventricular depolarization due to atrial lead malposition in the right ventricular outflow tract.
Answers are grounded in this paper and cite their sources.
May prompt lead position review in unexplained pacemaker-mediated tachycardia; hypothesis-generating for sensing anomalies in device troubleshooting.
A 44-year-old female had Chagas' disease with right bundle branch block, right axis deviation, and ventricular ectopic beats. An invasive evaluation showed a prolonged HV interval of 100 ms. A Medtronic Versatrax 7000 was implanted, after which a pacemaker-mediated tachycardia was observed. The esophageal lead ruled out the possibility of an atrial triggered tachycardia. The diagnosis was pacemaker-mediated tachycardia due to abnormal sensing of the terminal forces of the ventricular activation because of malposition of the atrial lead in the right ventricular out-flow tract.