Key result
A universal AV sequential pacemaker programmed in DVI mode induced pacemaker tachycardia at 150 bpm provoked by pectoral muscle myopotentials in a 67-year-old male.
Case Report (n=1)
This case highlights that myopotential interference can induce pacemaker tachycardia even in pulse generators programmed in the DVI mode.
Answers are grounded in this paper and cite their sources.
Myopotential-triggered tachycardia may occur despite DVI programming; this case leaves open generalizability and mitigation strategies.
A 67-year-old male, suffering from ventricular tachycardia unresponsive to drug therapy, received a universal AV sequential pacemaker (DDD,M). The pacemaker was programmed in the DVI mode, pacing rate 100 bpm, AV interval 250 ms. After implantation, the patient experiences two episodes of tachycardia that proved to be pacemaker tachycardia with a rate of 150 bpm. The first period was self-terminating, and the second had to be stopped by reprogramming the pulse generator. Pacemaker tachycardia could easily be provoked by instructing the patient to contract the pectoral muscle adjacent to the pulse generator. To our knowledge, this is the first report of pacemaker tachycardia provoked by myopotentials in a pulse generator programmed in the DVI mode.
No takes yet. Share an insight, caveat, or question.
GELDER et al. (1984) conducted a case report in Ventricular tachycardia unresponsive to drug therapy (n=1). Universal AV sequential pacemaker (DDD,M) programmed in DVI mode was evaluated on Pacemaker tachycardia provoked by myopotentials. A universal AV sequential pacemaker programmed in DVI mode induced pacemaker tachycardia at 150 bpm provoked by pectoral muscle myopotentials in a 67-year-old male.
Synapse has enriched 10 related papers. Consider them for comparative context: