Key result
Detergent LDL-C assays cross-react with ~51% of lipoprotein-X cholesterol, driving cholestatic measurement bias.
Observational (n=93)
Answers are grounded in this paper and cite their sources.
BACKGROUND: The purpose of this study was to evaluate the performance of two homogeneous assays for LDL-cholesterol (LDL-C), a polyethylene/cyclodextrin (PC) assay and a detergent (D) assay, which are based on different principles, in cholestatic serum. METHODS: We compared serum LDL-C concentrations determined by the two assays for healthy normolipidemic subjects (n = 42) and cholestatic patients (n = 51). LDL-C concentrations obtained with the homogeneous assays were also compared with those obtained by HPLC for patients' sera. In the interference study, conjugated bile acids were added to normal serum, and their effects on the two assays were examined. The effects of lipoprotein-X (LP-X), intermediate-density lipoprotein (IDL), and apolipoprotein (apo) E-rich HDL on the LDL-C assays were also investigated by adding these lipoproteins to normal serum. RESULTS: The LDL-C concentrations obtained with the D assay were higher than those obtained with the PC assay in the serum with high LP-X. The bias for LDL-C vs LP-X in cholestatic serum correlated with LP-X concentration (r = 0.582; P: <0.0001; n = 51). In the interference study, no effect of bile acids on the LDL-C assays was observed. However, the D assay measured 51.0% of the cholesterol in LP-X, whereas no reactivity was observed for LP-X in the PC assay. In addition, the D assay and the PC assay measured IDL-cholesterol at 31.2% and 52.4%, respectively, and measured apo E-rich HDL-C at 7.6% and 17.8%, respectively. CONCLUSIONS: Although both homogeneous LDL-C assays are suitable for most cases, the present study showed that each homogeneous assay has a different limitation for cholestatic serum with gross alterations in lipoproteins.
No takes yet. Share an insight, caveat, or question.
Hua et al. (2000) conducted an observational in Cholestasis with lipoprotein-X-positive serum (n=93). Detergent (D) homogeneous LDL-cholesterol assay vs. Polyethylene/cyclodextrin (PC) homogeneous LDL-cholesterol assay and HPLC was evaluated on Reactivity to lipoprotein-X (LP-X) cholesterol and assay bias in cholestatic serum (r = 0.582, p=<0.0001). The detergent LDL-C assay measured 51.0% of lipoprotein-X cholesterol compared with no reactivity in the polyethylene/cyclodextrin assay, correlating with LP-X bias.
Synapse has enriched 10 related papers. Consider them for comparative context: