BACKGROUND: HPV assay (Onclarity) on the BD Viper™ LT system (BD Diagnostics, Sparks, MD), detects E6/E7 DNA from 13 high-risk HPV genotypes and HPV66. We compared the analytical and clinical performance of the Onclarity Assay to that of Hybrid Capture 2 and LINEAR ARRAY using adjudicated histological outcomes from Danish women referred for colposcopy. METHODS: 276 women from Copenhagen, Denmark were referred for colposcopy with abnormal cytology and/or a positive HPV test. Two samples for HPV analysis were taken in BD SurePath™ and in the BD cervical brush diluent (CBD) media. ClinicalTrial gov. identifier: NCT01671462, Ethical Approval: H-4-2012-070. RESULTS: Histology was normal in 84 (31%) women, 70 (26%) had CIN1, 47 (17%) CIN2, and 68 (25%) had CIN3. The Onclarity assay detected 67 out of 68 (99%) ≥CIN3 and 113/115 (98%) ≥CIN2. The specificities for <CIN2 were 21%, 17%, and 22%, for HC2, Onclarity and LA, respectively. CONCLUSION: Overall, the Onclarity HPV assay performed well on SurePath LBC and CBD media, with clinical sensitivity and specificity matching those of HC2 and LA.
No takes yet. Share an insight, caveat, or question.
Ejegod et al. (2016) studied this question.