An improved network scheme has been devised to solve the linear simultaneous equations associated with superposed pulse-height spectra encountered in x-ray emission microanalysis by Hall and Mitra in 1969. The scheme has been developed specifically to retrieve x-ray characteristic line intensities from the pulse-height distribution produced by a proportional counter or solid-state detector by detection of the x-rays excited from a sample. Dolby's network scheme of 1961 has been improved so that the network coefficients may now be determined non-iteratively in the general n -element case. This makes calibration faster and simpler to the point where it may be automated if desired. An apparatus according to the new design can be linked readily to a commercial microanalyser. The network alignment procedure and the solution of the simultaneous equations in terms of the general n -element case is discussed.