Answers are grounded in this paper and cite their sources.
PURPOSE: To measure the adhesiveness of the AcrySof intraocular lens (IOL) to a collagen film to ascertain its capacity to adhere to the lens capsule. SETTING: Haibara General Hospital, Haibara, and Hamamatsu University School of Medicine, Hamamatsu, Shizuoka, Japan. METHODS: Each of five AcrySof IOLs was attached to a collagen film, slightly pressed in intraocular irrigating solution, and then pulled up and off the film with a DuNouy's surface balance. The force recorded by the surface balance at the moment it detached from the film was defined as the lens' adhesiveness. A control experiment was performed using five conventional poly(methyl methacrylate) (PMMA) IOLs. RESULTS: Mean adhesiveness was 2.76 gram-weight (gw) +/- 0.55 (SD) for AcrySof and 0.81 +/- 0.20 gw for PMMA. The difference was statistically significant (P < .0001, t-test). CONCLUSION: AcrySof may have a strong tendency to adhere to the lens capsule, contributing to posterior and anterior capsule clarity and preventing lens decentration in vivo.
No takes yet. Share an insight, caveat, or question.
Nagata et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: