Key result
Beating atrial septal defect closure monitored by real-time 3D echocardiography was feasible, with RT3DE ASD area measurements (82.5 mm2) closely correlating with direct measurements (81.6 mm2).
Why the study?
Is beating atrial septal defect closure monitored by epicardial real-time 3D echocardiography without cardiopulmonary bypass feasible in a canine model?
Is beating atrial septal defect closure monitored by epicardial real-time 3D echocardiography without cardiopulmonary bypass feasible in a canine model?
Answers are grounded in this paper and cite their sources.
Beating ASD closure without cardiopulmonary bypass is feasible in a canine model when guided by epicardial real-time 3D echocardiography and specialized suturing devices.
BACKGROUND: We assessed the feasibility of beating atrial septal defect (ASD) closure monitored by real-time 3D echocardiography (RT3DE). METHODS AND RESULTS: RT3DE was developed with prototype ultrasound equipment consisting of a high-speed 3D rendering unit with a frame rate of 5 to 10 frames/s. We also developed a prototype semiautomatic suture device and suture cutting system. In the experiment, 12 mongrel dogs were anesthetized, and after median sternotomy, the echo probe was applied directly to the surface of the right atrium. Three surgical maneuvers (balloon atrial septectomy, enlargement of the ASD, and ASD closure) were performed through the atrial port inserted into the right atrial appendage. The heart was then excised, and the area of the ASD measured by RT3DE was compared with its area measured directly. The ASD was successfully closed in all experimental animals except the first 2. Examination of the excised heart showed that none of the sutures were loose. The mean area of the ASD was 82.5+/-38.6 mm(2) when measured by RT3DE and 81.6+/-38.2 mm(2) when measured directly, and there was a significant correlation between the areas measured by RT3DE and those measured directly (echo measurements=1.007xdirect measurements+0.337; P<0.0001). A Bland-Altman analysis revealed close agreement between the results obtained by the 2 methods (7.807 mm(2) upper and -6.024 mm(2) lower limit of agreement). CONCLUSIONS: Introduction of RT3DE, a semiautomatic suture device, and a suture cutting system made beating ASD closure without cardiopulmonary bypass possible.
No takes yet. Share an insight, caveat, or question.
Suematsu et al. (2003) studied Atrial septal defect (n=12). Beating atrial septal defect closure monitored by real-time 3D echocardiography vs. Direct measurement was evaluated on Area of the ASD measured by RT3DE compared with direct measurement (p=<0.0001). Beating atrial septal defect closure monitored by real-time 3D echocardiography was feasible, with RT3DE ASD area measurements (82.5 mm2) closely correlating with direct measurements (81.6 mm2).