Key result
An implantable accelerometer-based heart-monitoring device prototype demonstrated sufficient stability in heart tissue and met IEC60601 leakage current requirements.
A new compact implantable accelerometer-based device for ischemia monitoring demonstrated adequate stability and electrical safety in preliminary testing.
Answers are grounded in this paper and cite their sources.
Animal data support further device development; leaves open translation to human ischemia monitoring.
This paper reports recent results from an ongoing effort to develop an implantable accelerometer-based heart-monitoring device for ischemia monitoring. The latest device prototype utilizes a new and more compact accelerometer (1.2 × 1.5 × 0.8 mm3), a prototype device from Bosch SensorTec, Reutlingen, Germany. This paper presents the fabrication and testing of the device, including an explorative study of the effect of the capsule shape on the stability of the implanted device in the heart tissue. The stability study indicated sufficient stability of the device and a higher resistance to retraction for one of the capsule designs. The device was able to carry out acceleration monitoring and it meets the leakage current requirements of the IEC60601 standard.
No takes yet. Share an insight, caveat, or question.
Tjuļkins et al. (2016) studied Ischemia. Implantable accelerometer-based heart-monitoring device was evaluated on Device stability and acceleration monitoring capability. An implantable accelerometer-based heart-monitoring device prototype demonstrated sufficient stability in heart tissue and met IEC60601 leakage current requirements.