Key result
An improved, smaller MEMS accelerometer-based heart monitoring device with simplified implantation demonstrated leakage current recordings well below standard limits in an animal experiment.
A newly developed MEMS accelerometer-based heart monitoring device demonstrates simplified implantation and safe leakage current levels in an animal model.
Answers are grounded in this paper and cite their sources.
Should not yet change clinical heart monitoring; leaves open translation of this animal-validated system to humans.
An accelerometer-based heart monitoring system has been developed for real-time evaluation of heart wall movement. In this paper, assembly and fabrication of an improved device is presented along with system characterization and test data from an animal experiment. The new device is smaller and has simplified the implantation procedure compared to earlier prototypes. Leakage current recordings were well below those set by the corresponding standards.
No takes yet. Share an insight, caveat, or question.
Tjuļkins et al. (2014) studied Heart wall movement monitoring. MEMS accelerometer-based heart monitoring device vs. Earlier prototypes was evaluated on Leakage current recordings. An improved, smaller MEMS accelerometer-based heart monitoring device with simplified implantation demonstrated leakage current recordings well below standard limits in an animal experiment.