Key result
Point-of-care coagulation assays showed limited clinical decision agreement with standard laboratory assays (e.g., 47% for warfarin plus heparin, 73% for warfarin) and are not interchangeable.
Why the study?
Do point-of-care assays provide comparable INR and aPTT measurements to standard laboratory assays in patients receiving various anticoagulant therapies?
Observational (n=150)
No
Do point-of-care assays provide comparable INR and aPTT measurements to standard laboratory assays in patients receiving various anticoagulant therapies?
Point-of-care coagulation assays show limited agreement with standard laboratory assays across various anticoagulant regimens and should not be considered interchangeable.
Supports continued use of laboratory assays for anticoagulant decisions; leaves open whether refined point-of-care tests merit prospective validation.
STUDY OBJECTIVE: To compare point-of-care and standard hospital laboratory assays for monitoring patients receiving single or combination anticoagulant regimens. DESIGN: Prospective analysis. SETTING: Nursing units and clinics at a large, community hospital. PATIENTS: One hundred fifty patients receiving anticoagulants for cardiac, vascular, orthopedic, or cancer indications. Thirty patients were enrolled into each treatment group: warfarin, enoxaparin, heparin, warfarin plus enoxaparin, and warfarin plus heparin. INTERVENTION: Capillary and venous blood samples were collected once in each patient for simultaneous measurement of international normalized ratio (INR) and activated partial thromboplastin time (aPTT) by both assays. MEASUREMENTS AND MAIN RESULTS: Mean differences in paired INR and paired aPTT by point-of-care and standard assays were small, but 95% confidence intervals were wide. The INR differences were greater for the warfarin plus heparin group than for the warfarin group or warfarin plus enoxaparin group; clinical decision agreement was 47% for warfarin plus heparin, 73% for warfarin, and 93% for warfarin plus enoxaparin. The aPTT difference was greater for the warfarin plus heparin than for the heparin group; however, clinical decision agreement, 67% and 70%, respectively, was similar. CONCLUSIONS: Point-of-care methods showed limited agreement with standard hospital laboratory assays of coagulation for all treatment groups. For INR values, significantly greater disagreement was noted between the assay methods for the warfarin plus heparin group compared with the warfarin group, but the agreement was similar for the warfarin and warfarin plus enoxaparin groups. Our data indicate that the point-of-care assays should not be considered interchangeable with standard laboratory assays.
No takes yet. Share an insight, caveat, or question.
Reiss et al. (2002) conducted an observational in Receiving anticoagulants for cardiac, vascular, orthopedic, or cancer indications (n=150). Point-of-care coagulation assays vs. Standard hospital laboratory assays was evaluated on Mean differences and clinical decision agreement in paired INR and aPTT. Point-of-care coagulation assays showed limited clinical decision agreement with standard laboratory assays (e.g., 47% for warfarin plus heparin, 73% for warfarin) and are not interchangeable.