Use of early solvent vapor exits for concurrent eluent evaporation with the loop‐type interface has two purposes: protection of the GC detectors from large amounts of solvent vapors and more efficient discharge of the vapors to accelerate eluent evaporation and help avoiding broad solvent peaks. Use of a retaining pre‐column after the uncoated pre‐column can rule out losses of solute materials that form sharp peaks.
No takes yet. Share an insight, caveat, or question.
Grob et al. (1989) studied this question.