Key result
A nonradioactive rubidium efflux assay using flame atomic absorption spectrometry yielded comparable EC50 (1-4 microM) and IC50 values for BKCa channel modulators as the radioactive 86Rb+ assay.
Flame atomic absorption spectrometry provides a viable, nonradioactive alternative to radioactive rubidium efflux assays for high-throughput screening of potassium channel modulators.
Although techniques such as (86)Rb(+) flux provide a sensitive measure of K(+) channel activity, the relatively short half-life and high-energy emission, together with the quantities of radioactive material generated, hinder the usefulness of flux-based formats in high throughput screening efforts. This study elaborates on the utilization of flame atomic absorption spectrometry (AAS) techniques for a nonradioactive rubidium efflux assay for large conductance Ca(2+)-activated K(+) channels (BK(Ca)) channels. Utilizing human embryonic kidney (HEK293) cells expressing the BK(Ca) alpha subunit, a 96-well cell-based nonradioactive rubidium efflux screen for channel openers and inhibitors was established. Known BK(Ca) channel openers, including NS1608, NS1619, and NS-8, activated rubidium efflux with EC(50) values ranging from 1 to 4 microM in both radioactive and nonradioactive efflux formats. Compounds such as iberiotoxin, paxilline, and charybdotoxin inhibited rubidium efflux responses evoked by the BK(Ca) channel opener NS1608 in both radioactive and nonradioactive efflux formats. The IC(50) values of the inhibitors in AAS format were comparable to those derived from (86)Rb(+) efflux assays. The present studies show that the pharmacological profiles of BK(Ca) channels assessed by AAS compare well with those obtained using the (86)Rb(+) efflux assay, and support the utility of nonradioactive efflux format for higher throughput screening campaigns for novel K(+) channel modulators.
No takes yet. Share an insight, caveat, or question.
Groebe et al. (2003) studied this question. Flame atomic absorption spectrometry (AAS) nonradioactive rubidium efflux assay vs. Radioactive (86)Rb(+) efflux assay was evaluated on Pharmacological profiles (EC50 and IC50 values) of BKCa channel openers and inhibitors. A nonradioactive rubidium efflux assay using flame atomic absorption spectrometry yielded comparable EC50 (1-4 microM) and IC50 values for BKCa channel modulators as the radioactive 86Rb+ assay.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: