Key result
Nonradioactive rubidium efflux assay yields ~0.60 correlation with conventional electrophysiology, outperforming membrane potential assays.
Why the study?
The lack of direct high throughput assays and the low throughput of conventional electrophysiology have impeded rapid screening of KCNQ/M-channel modulators.
Effect estimate: r = 0.60
The nonradioactive Rb+ efflux assay using ICR 8000 provides a robust, higher-throughput alternative to conventional electrophysiology for screening KCNQ2/3 channel modulators.
Rubidium efflux assay may improve KCNQ2/3 screening correlation; leaves open adoption over electrophysiology in neurological drug development.
M-channels (M-current), encoded by KCNQ2/3 K(+) channel genes, have emerged as novel drug targets for a number of neurological disorders. The lack of direct high throughput assays combined with the low throughput of conventional electrophysiology (EP) has impeded rapid screening and evaluation of K(+)-channel modulators. Development of a sensitive and efficient assay for the direct measurement of M-current activity is critical for identifying novel M-channel modulators and subsequent investigation of their therapeutic potential. Using a stable CHO cell line expressing rat KCNQ2/3 K(+) channels confirmed by EP, we have developed and validated a nonradioactive rubidium (Rb(+)) efflux assay in a 96-well plate format. The Rb(+) efflux assay directly measures the activity of functional channels by atomic absorption spectroscopy using the automated Ion Channel Reader (ICR) 8000. The stimulated Rb(+) efflux from KCNQ2/3-expressing cells was blocked by the channel blockers XE991 and linopirdine with IC(50) values of 0.15 microM and 1.3 microM, respectively. Twelve compounds identified as KCNQ2/3 openers were further assessed in this assay, and their EC(50) values were compared with those obtained with EP. A higher positive correlation coefficient between these two assays (r = 0.60) was observed than that between FlexStation membrane potential and EP assays (r = 0.23). To simplify the assay and increase the throughput, we demonstrate that EC(50) values obtained by measuring Rb(+) levels in the supernatant are as robust and consistent as those obtained from the ratio of Rb(+) in supernatant/lysate. By measuring the supernatant only, the throughput of ICR8000 in an eight-point titration is estimated to be 40 compounds per day, which is suitable for a secondary confirmation assay.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2004) studied Neurological disorders (KCNQ2/3 K+ channel targets). Rubidium (Rb+) efflux assay using Ion Channel Reader 8000 vs. Conventional electrophysiology (EP) and FlexStation membrane potential assays was evaluated on Correlation of EC50 values between assays (r = 0.60). A nonradioactive rubidium efflux assay for KCNQ2/3 K+ channels demonstrated a higher positive correlation with conventional electrophysiology (r = 0.60) than membrane potential assays.