Intravascular ultrasound is useful for identifying mechanical causes of balloon rupture during coronary interventions, such as protruding calcified ridges.
Answers are grounded in this paper and cite their sources.
IVUS may identify occult causes of recurrent balloon rupture; this case report leaves open routine adoption pending prospective data.
We describe a patient in whom balloon rupture occurred three times during inflation in a stent with restenosis in the left anterior descending artery. The cause of rupture was detected by intravascular ultrasound: a calcified ridge that protruded between the stent struts in the distal stent body. Cathet. Cardiovasc. Intervent. 50:212–214, 2000.
No takes yet. Share an insight, caveat, or question.
Gilutz et al. (2000) studied this question.