Why the study?
Do second-generation carotid mesh stents reduce stroke, death, myocardial infarction, and in-stent restenosis compared to single-layer stents in patients undergoing carotid artery stenting?
Do second-generation carotid mesh stents reduce stroke, death, myocardial infarction, and in-stent restenosis compared to single-layer stents in patients undergoing carotid artery stenting?
Answers are grounded in this paper and cite their sources.
Second-generation carotid mesh stents do not exhibit a uniform class effect; while CGuard reduces both clinical events and restenosis, the dual-braided Casper/Roadsaver stent is associated with delayed plaque prolapse, layer separation, and higher in-stent restenosis.
Whenever feasible, strokes should be prevented. 1Carotid-related strokes, presenting as isolated extracranial or "tandem" (extracranial plus intracranial) lesions, account for 20% to 30% of ischemic strokes 1 ; they are often large and disabling. 1,2Revascularization plays an important role in reducing carotid-related stroke incidence, particularly in increased-risk stenoses. 1istorically, carotid revascularization was best performed with surgery, but endovascular techniques have evolved significantly over the past 3 decades and are now an accepted alternative to endarterectomy. 1 Today, neuroprotected carotid artery stenting (CAS) employing second-generation stents has outcomes similar to surgery, with some stent designs likely superior to surgery. 3 Second-generation carotid stents (ie, antiembolic stents, dual-layer "mesh" stents) were developed to address the problem of incomplete plaque sequestration from the lumen, underlying a relative excess of 30-day minor strokes with single-layer stent CAS vs endarterectomy. 1,4o date, 3 carotid mesh stent designs have undergone clinical evaluation: Casper/Roadsaver (Terumo), Gore Carotid Stent (Gore), and CGuard MicroNet-covered stent (InspireMD).Casper/Roadsaver is a woven-wire, double metallic layer (each layer: braided) stent. 5The ~180-μm thick woven wires (stent frame) cover a woven mesh of 42-μm wires.The outer layer resembles a braided single-layer carotid stent (Wallstent); the inner metallic layer forms pores of ~0.15 to 0.20 mm 2 (Figure 1A). [5][6]6][7] The inner mesh is connected to the outer frame by ~45-μm thick wires. 5The dual-braided structure of the Casper/Roadsaver stent results in poor adaptation to curves/lumen irregularities. 5The stent mean thickness, determined by 4 wire layers, is ~1.7 mm (note ample space needing endothelization). 5This, taken together with a propensity for malapposition, may be prothrombogenic.Also, the final (postimplantation) stent length is difficult to predict (~20%-30% elongation). 5Advantages are the low-profile (5F), flexibility on delivery, and feasibility to resheath after partial release.The Gore Carotid Stent is composed of an open-cell (laser-cut) metallic frame and an (fixed on stent edges) ultra-thin Teflon outer layer with square-like perforations of ~0.25 mm 2 . 8Depending on the diameter, the delivery profile is 5F or 6F.
No takes yet. Share an insight, caveat, or question.
Musiałek et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: