Key result
Double-layer micromesh carotid stents show similar 30-day safety and outcomes to hybrid stents.
Why the study?
Periprocedural embolism risk differs between double-layer micromesh and hybrid carotid stent designs in high-grade carotid artery stenosis, but comparative safety and outcomes remain unclear.
Does a double-layer micromesh carotid stent system compared to a hybrid carotid stent system improve safety and outcomes in elective patients with high-grade carotid artery stenosis?
Cohort (n=67)
Open-label
No
Does a double-layer micromesh carotid stent system compared to a hybrid carotid stent system improve safety and outcomes in elective patients with high-grade carotid artery stenosis?
Double-layer micromesh and hybrid carotid stent systems demonstrate similar safety and 30-day outcome profiles in patients undergoing endovascular therapy for high-grade carotid artery stenosis.
Double-layer micromesh stents were associated with fewer periprocedural neurological events in elective CAS; leaves open whether this extends to randomized comparisons with single-layer designs.
Background: One of the major periprocedural risks of carotid artery stenting is embolism caused either by plaque debris or by local thrombus forming. Double-layer micromesh stent design has shown to lower the chance of debris embolism but might have a slightly higher risk of local thrombus forming. Thus, we compared two different stent designs regarding safety and outcome profile in elective patients with high-grade carotid artery stenosis using a self-expanding, double-layer micromesh carotid stent system (DLCS) or a self-expanding hybrid carotid stent system (HCS). Methods: A single-center, open-label, retrospective cohort study of 67 consecutive, elective patients with high-grade symptomatic and asymptomatic carotid stenosis was executed at a comprehensive stroke center. Outcome measures were reocclusion rate, periprocedural symptomatic ischemic events, as well as other periprocedural complications, and recurrent stroke and mortality at 30 days’ follow-up. Results: Thirty-two patients (24% women, median age 75 years (interquartile range (IQR) 71–80) were treated with DLCS, and 35 patients (29% women, median age 71 years (IQR 63–76) years) with HCS. In both groups, pretreatment carotid stenosis degree was similar (median NASCET of 80%). Successful deployment was achieved in all cases without technical failure, and both groups did not differ in reocclusion rates, recurrent stroke, and mortality within 30 days. Conclusions: DCLS and HCS revealed to have similar safety and outcome profile in elective patients with high-grade symptomatic as well as asymptomatic carotid artery stenosis.
No takes yet. Share an insight, caveat, or question.
Gruber et al. (2020) conducted a cohort in High-grade symptomatic and asymptomatic carotid artery stenosis (n=67). Double-layer micromesh carotid stent system (DLCS) vs. Hybrid carotid stent system (HCS) was evaluated on Reocclusion rate, periprocedural symptomatic ischemic events, other periprocedural complications, recurrent stroke, and mortality at 30 days. Double-layer micromesh and hybrid carotid stents demonstrated similar safety and outcome profiles, with no difference in reocclusion rates, recurrent stroke, and mortality within 30 days.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: