Answers are grounded in this paper and cite their sources.
Nickel has a number of adverse biological effects that have made the use of nickel in biomedical implants controversial. Yet information about the distribution of nickel in tissues around nickel-containing implants is scarce. The purpose of the current study was to use a laser ablation technique, combined with inductively coupled mass spectroscopy, to assess the spatial distribution of nickel around nickel-containing implants in vivo. Polyethylene, pure nickel wire, or a nickel-containing alloy (Ni-Cr) were implanted subcutaneously into rats for 7 days. The tissues were analyzed for Ni content and inflammation at 1-mm intervals up to 5 mm away from the implants. The sham surgery sites and the polyethylene caused mild to moderate inflammation 1-2 mm from the implant site with no detectable nickel in the tissue. The nickel wire caused severe inflammation up to 5 mm away from the implant site with necrosis for 1 mm around the implant. Nickel concentrations reached 48 microg/g near the implants, falling exponentially to undetectable levels at 3-4 mm from the implants. The Ni-Cr wire caused inflammation equivalent to polyethylene, with less than 4 microg/g of nickel present in the tissue for 1-2 mm around the implants. The current study showed that the laser-ablation technique was well suited for the analysis of soft tissues for metal-ion content, and that the nickel distribution in tissues correlated well with overt tissue inflammation.
No takes yet. Share an insight, caveat, or question.
Wataha et al. (2001) studied this question.