Why the study?
Does decreasing pump filling increase prosthetic heart valve cavitation in an in vivo calf LVAD model?
Population
2 calves implanted with Pennsylvania State University Left Ventricular Assist Devices using Björk-Shiley…
Comparison
Decreased pump filling by increasing beat rate vs Normal pump filling
Design
Preclinical
Follow-up
44 days
Authors
Loading...
Low pump filling may increase hemolysis via cavitation in LVAD models; extends preclinical data but leaves open human relevance.
Does decreasing pump filling increase prosthetic heart valve cavitation in an in vivo calf LVAD model?
In vivo measurement using high-fidelity pressure transducers confirms that prosthetic heart valve cavitation occurs and increases with decreased pump filling, transiently affecting plasma hemoglobin.
Zapanta et al. (1996) studied this question.