In an in vitro model of a mechanical heart valve, varying the partial pressure of CO2 did not significantly affect cavitation intensity.
Answers are grounded in this paper and cite their sources.
No PCO2 effect on cavitation in vitro; leaves open in vivo relevance for mechanical valve patients.
In previous studies, we investigated the cavitation phenomenon in a mechanical heart valve using an electro-hydraulic total artificial heart. With this system, a 50% glycerin solution kept at 37 degrees C was used as the working fluid. We reported that most of the cavitation bubbles were observed near the valve stop and were caused by the squeeze flow. However, in these studies, the effect of the partial pressure of CO(2) on the mechanical heart valve cavitation was neglected. In this study, in order to investigate the effect of the partial pressure of CO(2) on mechanical heart valve cavitation using an electro-hydraulic total artificial heart, we controlled the partial pressure of the CO(2) in vitro. A 25-mm Medtronic Hall valve was installed in the mitral position of an electro-hydraulic total artificial heart. In order to quantify the mechanical heart valve cavitation, we used a high-speed camera. Even though cavitation intensity slightly increased with increases in the PCO(2) at heart rates of 60, 70 and 100 bpm, throughout the experiment, there was no significant difference between the PCO(2) and cavitation intensity.