Synapse
⌘+K
Synapse
PulseExploreClubsResearchersJournals
Instagram
HomeClubsExplore
March 22, 2000Journal of Biomechanical Engineering

Observation and Quantification of Gas Bubble Formation on a Mechanical Heart Valve

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Does cavitation and gas content influence the formation of stable gas bubbles on mechanical heart valves?

Population

Mock circulatory loop employing a Medtronic-Hall pyrolytic carbon disk valve in the mitral position

Comparison

Varying cavitation intensity and carbon dioxide… vs Absence of cavitation; varying O2 concentration

Design

Preclinical

Authors

HLHsin‐Yi LinNational Taipei University of TechnologyBBBrian A. BianccucciPennsylvania State UniversitySDSteven DeutschLawrence Livermore National Laboratory

Discussion

Loading...

Member takes

Implication

In vitro findings link cavitation and CO2 to stable bubbles on mechanical valves; leaves open clinical relevance and preventive strategies.

Structured PICO

Does cavitation and gas content influence the formation of stable gas bubbles on mechanical heart valves?

P
Population
Mock circulatory loop employing a Medtronic-Hall pyrolytic carbon disk valve in the mitral position
I
Intervention
Varying cavitation intensity and carbon dioxide (CO2) concentration
C
Comparator
Absence of cavitation; varying O2 concentration
O
Outcome
Formation and volume of stable gas bubblessurrogate

In vitro modeling demonstrates that cavitation is an essential precursor to stable gas bubble formation on mechanical heart valves, driven primarily by CO2 concentration.

Cite This Study

Lin et al. (2000) studied this question.

synapsesocial.com/papers/6a6f66e15671bbf00bc290f3https://doi.org/10.1115/1.1287171
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1A Method for Real-Time In Vitro Observation of Cavitation on Prosthetic Heart Valves1994 · 34 citations
  2. 2Incidence and Frequency of Cerebral Embolic Signals in Patients With a Similar Bileaflet Mechanical Heart Valve1995 · 69 citations
  3. 3Prevalence and characteristics of intracranial microemboli signals in patients with different types of prosthetic cardiac valves.1994 · 124 citations
  4. 4Effect of Nitrous Oxide on Venous Air Embolism1966 · 120 citations