Key result
The Empatica E4 wristband showed poor reliability for measuring HRV (RMSSD median ICC = -0.05) over 5-minute intervals compared with a Holter monitor, though 24-hour agreement was satisfactory.
Why the study?
Although wristband devices claim to measure HRV, none had demonstrated reliability compared with the criterion-standard Holter monitor.
Does the Empatica E4 wristband reliably measure heart rate variability compared to a Holter monitor in ambulatory patients?
Observational (n=13)
Does the Empatica E4 wristband reliably measure heart rate variability compared to a Holter monitor in ambulatory patients?
The Empatica E4 wristband demonstrates poor reliability for short-term (≤5 minutes) heart rate variability measurements compared to a Holter monitor, though 24-hour aggregate measures show satisfactory agreement.
Short-term E4 HRV unreliable for clinical use; leaves open 24-hour aggregate validation versus Holter in ambulatory research.
OBJECTIVE: Heart rate variability (HRV) is a useful index of psychological and physiological stress. Although several wristband devices have purported to measure HRV, none have demonstrated reliability when compared with the criterion-standard Holter monitor. We evaluated the reliability of HRV readings from the Empatica E4 wristband compared with a Holter monitor over a 24-hour period of simultaneous monitoring. METHODS: Agreement between the monitors was assessed by examining correlations and intraclass correlations (ICCs) for fixed sets in 13 individuals in a treatment trial for posttraumatic stress disorder (4 women; mean [standard deviation] age = 51.92 [6.17] years). Agreement was calculated at 1-second and 5-minute intervals for interbeat intervals (IBIs) and for 5-minute intervals of the root mean square of successive differences between normal heartbeats (RMSSD) and standard deviation of all normal R-R intervals (SDNN). Agreement across the entire 24-hour observation period was also measured. Frequency-domain measures of HRV could not be calculated because of too much missing data from the E4. RESULTS: Although high interdevice correlations and ICCs were observed between the E4 and Holter monitors for IBIs at 1-second (median r = 0.88; median ICC = 0.87) and 5-minute (median r = 0.94; median ICC = 0.94) intervals, reliabilities for 5-minute RMSSD (median r = -0.09; median ICC = -0.05) and 5-minute SDNN (median r = 0.48; median ICC = 0.47) were poor. Agreement between the devices on 24-hour measures of HRV was satisfactory (IBI: r = 0.97, ICC = 0.97; RMSSD: r = 0.77, IBI = 0.76; SDNN: r = 0.92, IBI = 0.89). CONCLUSIONS: Findings suggest that the low reliability of Empatica E4 as compared with the Holter monitor does not justify its use in ambulatory research for the measurement of HRV over time periods of 5 minutes or less.
No takes yet. Share an insight, caveat, or question.
Voorhees et al. (2021) conducted an observational in posttraumatic stress disorder (n=13). Empatica E4 wristband vs. Holter monitor was evaluated on Agreement between monitors for HRV measures (IBI, RMSSD, SDNN) at 1-second, 5-minute, and 24-hour intervals. The Empatica E4 wristband showed poor reliability for measuring HRV (RMSSD median ICC = -0.05) over 5-minute intervals compared with a Holter monitor, though 24-hour agreement was satisfactory.