Key result
Automated enzymatic triglyceride assay delivers high precision and linearity up to ~/dL.
Answers are grounded in this paper and cite their sources.
We describe a completely automated enzymatic method for estimating serum triglycerides with a centrifugal analyzer (ENI-GEMSAEC). Accuracy and precision (CV, 2 to 4.8%) and the upper limit of linearity (400 mg/100 ml) are monitored by use of newly developed, clear, aqueous triolein standards. A modified computer program (FOCAL 8) is used for data manipulation and correcting absorbance to zero reaction time. Results cornpare well with those obtained by an automated fluorometric procedure and a manual enzymatic procedure.
No takes yet. Share an insight, caveat, or question.
Chong-Kit et al. (1974) studied this question. Automated all-enzymatic serum triglyceride method on GEMSAEC centrifugal analyzer vs. Automated fluorometric procedure and manual enzymatic procedure was evaluated on Analytical precision (coefficient of variation) and linearity. The automated enzymatic triglyceride assay on the GEMSAEC analyzer achieved a CV of 2-4.8% and linearity up to /100 ml, correlating well with fluorometric and manual methods.