Key result
Modern coronary sinus leads demonstrated excellent long-term stability, with only 7 lead dislocations and 9.7% of patients showing impedance >1000 Ohm during follow-up.
Cohort (n=193)
No
Modern coronary sinus leads for cardiac resynchronization therapy demonstrate excellent long-term stability, performance, and safety in a real-world cohort.
Supports modern CS lead durability in CRT; extends real-world data but should not yet change practice without RCTs.
BACKGROUND: Cardiac resynchronization therapy (CRT) has become an important pillar of contemporary heart failure therapy. The efficacy of CRT, however, critically relies on proper LV lead placement and performance, which is why data regarding the long-term performance of CS leads are of considerable interest. Available studies are limited by a restricted variety of lead vendors, earlier lead models and / or very short follow-up periods. In the current study, we therefore investigated the long-term performance of modern LV leads in a large "real world" cohort of patients undergoing CRT implantation. METHODS AND RESULTS: All 193 patients who had successfullyundergone CRT implantation at the University Hospital Zurich between September 2003 and January 2010 were included in the study. An overall stable course of stimulation energy was observed over time; neither ischemic etiology, lead configuration, or severely reduced EF had an influence on the evolution of energy thresholds over time. Interestingly, patients with a high energy threshold at baseline experienced a significant reduction during follow-up. In contrast, a significant drop in impedance was seen following implantation, followed by a steady course for the rest of the observation period. Only 15 patients (9.7%) showed an impedance > 1000 Ohm at any time during their follow-up. Seven lead dislocations were observed during follow up. CONCLUSION: The current comprehensive analysis of long-term performance of modern coronary sinus leads demonstrates excellent stability, performance and safety. These data may have important implications for physicians involved in biventricular pacemaker implantations and in the follow-up care of these patients.
No takes yet. Share an insight, caveat, or question.
Steffel et al. (2014) conducted a cohort in Heart failure requiring cardiac resynchronization therapy (n=193). Modern coronary sinus leads was evaluated on Long-term lead performance (stimulation energy, impedance, and dislocations). Modern coronary sinus leads demonstrated excellent long-term stability, with only 7 lead dislocations and 9.7% of patients showing impedance >1000 Ohm during follow-up.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: