Key result
During a median follow-up of 645 days after CRT device implantation, 7% of patients required a coronary sinus lead intervention, with first endovascular replacement successful in 86% of cases.
Why the study?
What is the long-term requirement for coronary sinus lead interventions and the effectiveness of endovascular replacement after CRT implantation?
Cohort (n=568)
No
What is the long-term requirement for coronary sinus lead interventions and the effectiveness of endovascular replacement after CRT implantation?
Answers are grounded in this paper and cite their sources.
Coronary sinus lead interventions are required in 7% of patients after CRT implantation, but endovascular replacement is highly successful in resolving these issues.
AIMS: The aim of this study was to assess the requirement for coronary sinus (CS) lead intervention after cardiac resynchronization therapy (CRT) and to evaluate the effectiveness of endovascular replacement. METHODS AND RESULTS: All patients receiving a CRT device with CS lead in the Leiden University Medical Center in the period from 1999 to 2007 were prospectively evaluated and followed. Five hundred and seventy-seven patients were successfully implanted with a CRT device. Nine (1.6%) patients were lost to follow-up. The remaining 568 patients were included in the analysis. During a median follow-up time of 645 days (inter-quartile range, 260-1148), 7% of the patients required a CS lead intervention. Cause of the intervention was an elevated threshold (n = 13), loss of capture (n = 20), or intractable phrenic nerve stimulation (n = 6). Fifteen patients (38%) required a CS lead intervention before first scheduled follow-up (2 months after implantation). Thirteen patients (33%) warranted a CS lead intervention more than 6 months after implantation. The first endovascular replacement was successful in 86% (32 of 37), whereas a second endovascular approach failed in 66% (2 of 3). CONCLUSION: The long-term requirement for CS lead interventions is 7%. Endovascular repositioning or replacement is successful in the majority of cases.
No takes yet. Share an insight, caveat, or question.
Borleffs et al. (2009) conducted a cohort in Cardiac resynchronization therapy (n=568). Cardiac resynchronization therapy with coronary sinus lead was evaluated on Requirement for coronary sinus lead intervention. During a median follow-up of 645 days after CRT device implantation, 7% of patients required a coronary sinus lead intervention, with first endovascular replacement successful in 86% of cases.