This is the authors' abstract. We don't add key points for this paper.
We present a quantitative discussion of the acoustic transmission line theory pertaining to experimental results from a resonant photoacoustic cell excited in its first longitudinal mode. Window absorption is optimally suppressed by buffer volumes and tunable air columns. The acoustic behavior of an ultrasensitive one inch condenser microphone is quantitatively described. A small and sensitive photoacoustic cell has been developed for intracavity use in a CO2 waveguide laser permitting measurements of ethylene down to 6 pptv (long term stability 20 pptv) with a time response of 2 s at a trace gas flow of 6 1/h. To demonstrate the fast time response within a biological application the instant ethylene release of a single tomato is measured.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Bijnen et al. (1996) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: