Answers are grounded in this paper and cite their sources.
This study aimed to assess intra- and inter-visit reliability of a novel EasyForce dynamometer for assessment of maximal strength in knee extension and flexion, and hip abduction and adduction tasks. Young healthy participants (n = 50; 25 males, 25 females; age: 23.4 ± 2.1 years) were measured for maximal joint torque with EasyForce dynamometer and rigid isometric dynamometers with external fixation, in two-sessions (5–10 days apart). The intra-visit reliability was high (ICC) = 0.84–0.95 knee; 0.90–0.94 hip), with absolute reliability sometimes exceeding the acceptable threshold (CV = 6.9–10.6% knee; 7.8–11.8% hip). Inter-visit reliability was excellent for all tasks (ICC = 0.95–0.97 knee; 0.95–0.96 hip). Although the EasyForce values were different compared to rigid dynamometers, the correlations between the devices were high (r = 0.64–0.75 knee; 0.70–0.75 hip). In conclusion, the EasyForce dynamometer provides reliable means of assessing knee (extension and flexion) and hip (abduction and adduction) muscle strength.
No takes yet. Share an insight, caveat, or question.
Kozinc et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: