Why the study?
Is radial pulse wave analysis repeatable for measuring central aortic blood pressures in ambulant patients?
Is radial pulse wave analysis repeatable for measuring central aortic blood pressures in ambulant patients?
Non-invasive assessment of central aortic pressures using radial pulse wave analysis is highly repeatable in ambulant patients, even when performed by relatively inexperienced staff.
Answers are grounded in this paper and cite their sources.
Supports radial PWA for central BP assessment by varied staff; leaves open whether it improves outcomes versus brachial measurements.
OBJECTIVE: To estimate the repeatability of radial pulse wave analysis (PWA) in measuring central systolic and diastolic blood pressures (cSBP/cDBP), pulse pressure (cPP), augmentation pressure (cAP) and pulse pressure amplification (PPA). METHODS: After 15 min supine rest, 20 ambulant patients (aged 27-82 years; four female) underwent four SphygmoCor PWA measurements on a single occasion. Two nurses independently undertook two measurements in alternate order, blind to their colleague's measurements. Analysis was by Bland-Altman limits of agreement (LOA). RESULTS: Heart rate and brachial blood pressure (BP) were stable during assessment. Based on the average of two PWA measurements between-observer differences (LOA, mean difference +/- 2SD) were small (cSBP 1.5 +/- 10.9 mmHg; cDBP 0.4 +/- 5.2 mmHg; cAP 0.5 +/- 4.5 mmHg; cPP 1.1 +/- 10.5 mmHg; PPA -0.5% +/- 5.6%). Between-observer differences were much greater for single/initial PWA measurement (cSBP 3.6 +/- 15.9 mmHg; cDBP 2.8 +/- 8.8 mmHg; cAP 0.7 +/- 5.8 mmHg; cPP 0.8 +/- 13.6 mmHg; PPA -1.2 +/- 9.4%). Within-observer LOA were very similar for both nurse A (cSBP -4.2 +/- 14.1 mmHg; cDBP -4.6 +/- 13.1 mmHg; cAP -0.4 +/- 4.4 mmHg; cPP 0.5 +/- 11.0 mmHg; PPA 0.7% +/- 9.0%) and nurse B (cSBP 0.0 +/- 12.1 mmHg; cDBP 0.2 +/- 8.5 mmHg; cAP -0.1 +/- 4.4 mmHg; cPP -0.2 +/- 11.9 mmHg; PPA -0.7% +/- 10.6%). CONCLUSION: Non-invasive assessment of central aortic pressures using PWA on a single occasion is highly repeatable in ambulant patients even when used by relatively inexperienced staff.
No takes yet. Share an insight, caveat, or question.
Coch et al. (2007) studied this question.