Captopril, a thiol-containing angiotensin-converting enzyme (ACE) inhibitor, is known for its limited stability in biological matrices due to oxidative degradation. This study introduces a validated liquid chromatography–tandem mass spectrometry (LC–MS/MS) method for the quantification of captopril in human plasma, with stabilization achieved through the addition of sodium metabisulphite. Chromatographic separation was carried out using a Zorbax CN (Cyano) column measuring 75 × 4.6 mm with a particle size of 3.5 µm. The mobile phase consisted of methanol and 0.1% formic acid in a 70:30 (v/v) ratio. Detection utilized negative electrospray ionization mode, monitoring multiple reaction monitoring (MRM) transitions of 216.1 → 114.2 for captopril and 219.3 → 117.3 for the internal standard, Captopril D3. Sample preparation involved solid-phase extraction using MCX cartridges. The method was validated in accordance with ICH M10 guidelines. It demonstrated excellent linearity across the concentration range of 5.00 to 799.56 ng/mL, with high sensitivity, accuracy ranging from 99.14 to 101.69%, and precision with a coefficient of variation not exceeding 4.99%. The incorporation of sodium metabisulphite significantly enhanced the stability of captopril in plasma samples. This validated method is suitable for use in pharmacokinetic studies, bioequivalence assessments, and therapeutic drug monitoring.