Answers are grounded in this paper and cite their sources.
An optimal control algorithm for IABP deflation time relative to the R wave improves optimization speed during heart rate variations in animal models.
An optimal control algorithm was adapted to identify and track the optimal deflation time of the intraaortic balloon pump (IABP). Routines for handling physiologically imposed constraints were added to the algorithm which was implemented in a computer-controlled system. The system was designed to provide real time optimization for the clinical setting. The controller continuously maximizes a performance index while observing the constraints. The index is composed of clinically available hemodynamic variables which indicate changes in myocardial energy balance. Proper values for the algorithm parameters were determined and the system was tested in animal experiments. The results indicate that controlling deflation time relative to the R wave, which precedes the next ejection phase, reduces the time required for optimization when the heart rate varies.
No takes yet. Share an insight, caveat, or question.
Barnea et al. (1992) studied this question.