Key result
Exposure to external electric fields caused up to a 200% variation in interference voltage at the input of unipolar cardiac pacemakers due to individual anatomical and physiological factors.
Why the study?
How do anatomical and physiological parameters influence the interference voltage of unipolar cardiac pacemakers in low-frequency electric fields?
Observational (n=15)
How do anatomical and physiological parameters influence the interference voltage of unipolar cardiac pacemakers in low-frequency electric fields?
Answers are grounded in this paper and cite their sources.
Individual anatomical and physiological factors can cause up to a 200% variation in interference voltage for unipolar pacemakers in electric fields, highlighting the need for individualized safety considerations.
The problem of electromagnetic interference of electronic implants such as cardiac pacemakers has been well known for many years. An increasing number of field sources in everyday life and occupational environment leads unavoidably to an increased risk for patients with electronic implants. However, no obligatory national or international safety regulations exist for the protection of this patient group. The aim of this study is to find out the anatomical and physiological worst-case conditions for patients with an implanted pacemaker adjusted to unipolar sensing in external time-varying electric fields. The results of this study with 15 volunteers show that, in electric fields, variation of the interference voltage at the input of a cardiac pacemaker adds up to 200% only because of individual factors. These factors should be considered in human studies and in the setting of safety regulations.
No takes yet. Share an insight, caveat, or question.
Joosten et al. (2009) conducted an observational in Implanted pacemaker adjusted to unipolar sensing (n=15). External time-varying electric fields was evaluated on Interference voltage at the input of a cardiac pacemaker. Exposure to external electric fields caused up to a 200% variation in interference voltage at the input of unipolar cardiac pacemakers due to individual anatomical and physiological factors.