Key result
Exposure to electric 50 Hz fields produced an interference voltage of 180 microVpp (real-case) and 400 microVpp (worst-case) per 1 kV m(-1), making pacemaker disturbance unlikely under practice-relevant conditions.
Why the study?
Does exposure to electric 50 Hz fields cause significant interference in implanted cardiac pacemakers?
Observational
Does exposure to electric 50 Hz fields cause significant interference in implanted cardiac pacemakers?
Electric 50 Hz fields beneath high voltage overhead lines are unlikely to disturb implanted cardiac pacemakers under practice-relevant conditions.
Answers are grounded in this paper and cite their sources.
Pacemaker patients may face low interference risk from 50 Hz fields; leaves open validation in larger clinical cohorts.
The aim of the study was to determine a 'worst-case' and a 'real-case' interference threshold for implanted cardiac pace-makers (CPM) in electric 50 Hz fields as they appear in high-voltage plants, e.g. beneath high voltage overhead lines. For this purpose the resulting electrical potential distribution within the thorax area of volunteers from an external homogeneous electrical 50 Hz field was measured. Different factors such as different body geometries as well as inspiration and expiration of the lung were considered. Measurements showed that 1 per 1 kV m(-1) unimpaired electrical field strength (RMS) an interference voltage of about 180 microVpp as real-case value an 400 microVpp as worst-case value would occur at the input of a unipolar ventricularly controlled, left pectorally implanted cardiac pacemaker. Therefore, it is possible under worst-case conditions but unlikely under practice-relevant conditions that an implanted cardiac pacemaker is disturbed by present electric 50 Hz fields beneath high voltage overhead lines.
No takes yet. Share an insight, caveat, or question.
Jan Scholten (2001) conducted an observational in Implanted cardiac pacemakers. Electric 50 Hz fields was evaluated on Interference voltage at the input of a unipolar ventricularly controlled, left pectorally implanted cardiac pacemaker. Exposure to electric 50 Hz fields produced an interference voltage of 180 microVpp (real-case) and 400 microVpp (worst-case) per 1 kV m(-1), making pacemaker disturbance unlikely under practice-relevant conditions.
Synapse has enriched 10 related papers. Consider them for comparative context: