There is a limited possibility for interference of implanted unipolar cardiac pacemakers by extremely low frequency magnetic fields in everyday life.
Answers are grounded in this paper and cite their sources.
Thresholds suggest limited everyday ELF interference for unipolar pacemakers; animal data leaves open human translation and clinical validation.
The effective induction loop area of implanted cardiac pacemaker (CPM) systems in magnetic fields was determined. The results were verified in a tank model placed in the centre of a Helmholtz-coil-arrangement. Both a left and a right pectorally implanted unipolar dual chamber CPM system were simulated. On this basis and with the results of benchmark-tests the interference thresholds for a collection of modern CPMs in extremely low frequency (ELF) magnetic fields were estimated. The investigations clearly showed that there are two loops, the CPM-lead-tissue-loop and the body loop, responsible for the magnitude of the disturbance voltage on the input of a cardiac pacemaker. The effective induction loop areas rangedfrom 100 to 221 cm2. For a left pectorally implanted, atrially controlled CPM system the interference thresholds for the magnetic induction lay between 16 and 552 micro T (RMS) for frequencies of the magneticfield between 10 and 250 Hz. Thus, there is a limited possibility for an interference of implanted CPM by ELF magnetic fields in everyday life.