Key result
60-Hz contact current interference thresholds for typical sensitivity settings range from 24 to 45 microA for unipolar pacemakers and 63 to 340 microA for bipolar pacemakers.
Why the study?
What are the 60-Hz contact current interference thresholds for implanted cardiac pacemakers?
Population
Millimeter-scale human body conductivity models with simulated implanted cardiac pacemakers
Design
Preclinical
Authors
Loading...
Animal thresholds suggest possible EMI risk from low 60-Hz currents; leaves open human applicability and need for prospective validation before practice change.
What are the 60-Hz contact current interference thresholds for implanted cardiac pacemakers?
Numerical modeling demonstrates that 60-Hz contact currents can cause interference with unipolar and bipolar pacemakers at thresholds of 24-45 microA and 63-340 microA, respectively.
Dawson et al. (2002) studied Implanted cardiac pacemakers. 60-Hz contact currents was evaluated on Interference thresholds for unipolar and bipolar pacemakers. 60-Hz contact current interference thresholds for typical sensitivity settings range from 24 to 45 microA for unipolar pacemakers and 63 to 340 microA for bipolar pacemakers.