Key result
Exposure to 50 Hz electromagnetic interference caused pacemaker malfunction at much lower current levels in unipolar mode compared to bipolar mode, which conferred a high degree of protection.
Why the study?
Does exposure to 50 Hz extraneous electrical interference cause pacemaker malfunction in patients with implanted dual chamber pacemakers?
Does exposure to 50 Hz extraneous electrical interference cause pacemaker malfunction in patients with implanted dual chamber pacemakers?
Answers are grounded in this paper and cite their sources.
Bipolar mode confers a high degree of protection from 50 Hz extraneous electrical interference, whereas unipolar mode is susceptible to clinically significant malfunction at lower current levels.
Twenty-two patients with dual chamber pacemakers with interchangeable lead configuration were exposed to 50 Hz electromagnetic interference. Current, at corporeal levels from 0-600 microA, was applied between electrodes on shoulders and feet using a bedside injection unit. Pacemaker behavior was monitored with telemetered event markers and intracardiac electrograms. In bipolar mode, noise reversion mode was induced in all except two Medtronic units at high (> 170 microA) levels of corporeal current. In the Intermedics, Siemens Pacesetter, and Telectronics models, onset of noise reversion mode was preceded by a window of inappropriate function, characterized by rate acceleration due to atrial malsensing, or pacemaker inhibition due to ventricular malsensing. In unipolar mode, pacemaker malfunction occurred at much lower current levels. Inappropriate behavior preceded the onset of protective noise reversion mode. During current injection, all pacemakers could be interrogated and reprogrammed, and intracardiac telemetry was reliably obtained except in two Medtronic units at high current levels. No pacemaker was reset by the electrical interference, and no cross-talk was seen. Use of bipolar mode confers a high degree of protection from extraneous electrical interference, but in unipolar mode pacemakers may be inhibited by small amounts of corporeal current, potentially encountered in every day life. The current injection unit allows safe, controllable, and quantifiable investigation of the effects of the electric field induced by a current on implanted pacemakers. Telemetry of annotated intracardiac signals during electromagnetic interference clarifies observations of pacemaker acceleration and inhibition.
No takes yet. Share an insight, caveat, or question.
ASTRIDGE et al. (1993) studied Patients with dual chamber pacemakers (n=22). 50 Hz electromagnetic interference was evaluated on Pacemaker behavior (noise reversion mode, inappropriate function). Exposure to 50 Hz electromagnetic interference caused pacemaker malfunction at much lower current levels in unipolar mode compared to bipolar mode, which conferred a high degree of protection.