Does injected 50 Hz alternating current affect the function of implanted unipolar VVI pacemakers?
Does injected 50 Hz alternating current affect the function of implanted unipolar VVI pacemakers?
Injected 50 Hz alternating current safely and reproducibly tests the sensitivity of implanted pacemakers to external interference, causing reversion to interference mode at 29-250 microA.
Answers are grounded in this paper and cite their sources.
May support individualized in vivo testing of pacemaker interference thresholds; leaves open clinical relevance to real-world electromagnetic exposure.
The effects of injected 50 Hz alternating current on the function of cardiac pacemakers has been observed in 18 patients with implanted unipolar VVI units. Current, in the range 0-600 microA was applied via electrodes attached to the patients' upper body and feet and fed from a specially designed current injection unit at the bedside. Most implanted pacemakers reverted to interference mode in the current range 29-250 microA. At current levels just below the reversion current all units developed irregular and inappropriate pacing. This current level was pacemaker dependent and varied in the range 27-246 microA. The total reversion current depended on the location of the injecting electrodes and on the patients' posture. The sensitivity of the units to injected interference was increased by deep inspiration. Temporary pacing catheters fitted to an additional ten patients were used to monitor the interference voltage which would be sensed by an implanted unit. This voltage was similarly dependent on patient posture and on deep respiration. Current injection has proved to be a safe, controllable and reproducible method of testing the sensitivity of implanted pacemakers to 50 Hz external interference.
No takes yet. Share an insight, caveat, or question.
Kaye et al. (1988) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: