This paper describes the development of a fiber-optic chemical sensor based on the principle of competitive-binding fluorescence immunoassay. Rabbit immunoglobln G (IgG) is covalently Immobilized on the distal sensing tip of a quartz optical fiber. The sensor is exposed to fluorescein Isothiocyanate (FITC) labeled and unlabeled anti-rabbit IgG. The 488-nm line of an argon-ion laser provides excitation of sensor-bound analyte. This results in fluorescence emission at the optical fiber‧s sensing tip. Sensor response is inversely proportional to the amount of unlabeled anti-IgG In the sample. Limits of detection (LOD) vary with Incubation time, sample size, and measurement conditions. For 10-μL samples, typical LOD are 25 fmol of unlabeled antibody In a 20-min Incubation period. These results Indicate that each fiber-optic fluoroimmunosensor can be constructed to perform a single sensitive, rapid, low-volume immunoassay, in in situ or benchtop applications.