Answers are grounded in this paper and cite their sources.
Plastic microchannels (4.5 cm long) fabricated from an etched glass master were tested for high-resolution single-stranded DNA analysis. Using replaceable denaturing linear polyacrylamide as sieving matrix, one-color separation of a fragment sizing standard with single-base resolution (R > 0.5) was achieved up to 275 bases. Two-color sizing analysis of four loci short tandem repeat (STR) allelic ladder (CSF1PO, TPOX, TH01, vWA) with single-base resolution (R = 0.62) on TH01 alleles 9.3 (198 bp) and 10 (199 bp) was demonstrated. An average standard deviation of +/- 0.06 bp and +/-0.11 bp in sizing 32 alleles of the CTTv ladder was attained between runs and between channels, respectively. Four-color sequencing separation of a terminator sequencing standard showed a base-calling accuracy of 99.1% out to 320 bases in 13 min.
No takes yet. Share an insight, caveat, or question.
Shi et al. (2003) studied this question.