Conventional and dedicated atrial overdrive pacing for the prevention of paroxysmal atrial fibrillation: the AFTherapy study

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Does conventional atrial overdrive pacing or dedicated preventive pacing algorithms reduce AF burden in patients with drug-refractory paroxysmal AF?

Population

372 patients with drug-refractory paroxysmal atrial fibrillation (AF)

Comparison

Dual-chamber pacing device delivering… vs Single rate pacing at 40 min-1 and conventional…

Design

RCT, Randomized to single rate or rate-responsive pacing with lower rates of…

Authors

A. John Camm
A. John CammElectrophysiology
NSNeil SulkeUniversity College LondonNENils EdvardssonElectrophysiology

Discussion

Loading...

Member takes

Overview

Preventive pacing algorithms confer no AF burden reduction; confirms conventional pacing at 70 bpm suffices without added algorithms in drug-refractory paroxysmal AF.

Structured PICO

Does conventional atrial overdrive pacing or dedicated preventive pacing algorithms reduce AF burden in patients with drug-refractory paroxysmal AF?

P
Population
372 patients with drug-refractory paroxysmal atrial fibrillation (AF)
I
Intervention
Dual-chamber pacing device delivering conventional overdrive pacing (single rate or rate-responsive pacing at lower rates of 70 or 85 min-1) and dedicated preventive pacing (lower rate of 70 min-1 with four preventive pacing algorithms)
C
Comparator
Single rate pacing at 40 min-1 (for conventional pacing phase) and conventional pacing at 70 min-1 without preventive algorithms (for preventive pacing phase)
O
Outcome
Atrial fibrillation (AF) burdensurrogate

Conventional atrial overdrive pacing and dedicated preventive pacing algorithms do not significantly reduce AF burden in patients with drug-refractory paroxysmal AF.

Limitations

  • A substantial amount of data was excluded from the analysis because of atrial-sensing artefacts identified in the device-captured diagnostics

Cite This Study

Camm et al. (2007) studied this question.

synapsesocial.com/papers/6a87d0076e1dca10b7bb5cf6https://doi.org/10.1093/europace/eum253
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Long‐term Prevention of Vagal Atrial Arrhythmias by Atrial Pacing at 90/Minute: Experience with 6 Cases1983 · 122 citations
  2. 2Far‐Field QRS Complex Sensing: Prevalence and Timing with Bipolar Atrial Leads2000 · 46 citations
  3. 3Atrial pacing for prevention of atrial fibrillation: assessment of simultaneously implemented algorithms2004 · 94 citations
  4. 4Efficacy and tolerability of continuous overdrive atrial pacing in atrial fibrillation2000 · 38 citations
  5. 5Effect of Right Atrial Overdrive Pacing in the Prevention of Symptomatic Paroxysmal Atrial Fibrillation:2003 · 43 citations