Why the study?
Is the implantation of an additional pace-sense lead a safe strategy for managing an isolated pace-sense problem in a defibrillation lead?
Is the implantation of an additional pace-sense lead a safe strategy for managing an isolated pace-sense problem in a defibrillation lead?
The placement of an additional pace-sense lead is a safe short-term management strategy for isolated pace-sense problems in defibrillation leads, avoiding immediate lead extraction.
Answers are grounded in this paper and cite their sources.
Additional lead implantation may avoid extraction for isolated pace-sense defects; leaves open long-term durability and need for controlled trials.
AIMS: The implantation of an additional pace-sense (P/S) lead is a standard treatment option in the management of an isolated pace-sense problem in a defibrillation (HV-P/S) lead. However, the safety of this management strategy is unclear. We performed a retrospective single-centre study to assess this. METHODS AND RESULTS: We studied all patients with an isolated P/S problem in an HV-P/S lead, treated with an additional P/S lead, in our institution. The need for further invasive intervention for a lead-related complication, or death during follow-up, was assessed. From 2000 to 2008, 45 patients were treated with an additional P/S lead. Mean follow-up was 78 +/- 38 months from original device implantation and 28 +/- 17 months following implantation of the additional lead. During follow-up, three patients required an invasive intervention for a lead-related problem. All were successfully treated with lead extraction and device re-implantation. There were five deaths. Following implant of an additional lead, cumulative survival from further lead defects after 6 months, 1, 2, and 3 years was 100, 100, 93, and 87%, respectively. CONCLUSION: In the treatment of an isolated P/S problem in an HV-P/S lead, the placement of an additional P/S lead is a safe management strategy, at least in the short term.
No takes yet. Share an insight, caveat, or question.
Scott et al. (2009) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: