Alanine dosemeters have been exposed to beams of 4, 8 and 16 MV p X-rays and 6, 8, 10, 12, 14, 17 and 20 MeV electrons from linear electron accelerators used for radiation therapy. The dose levels used were 100 Gy or below. The theoretical energy dependence of the dosemeters have been calculated both for photons and electrons. After correction for the theoretical energy dependence, the response of the dosemeters had a coefficient of variation of 0.4% around the average value. The average response was within 1% of the response of the alanine dosemeters exposed to a 60 Co gamma -ray source or a 137 Cs gamma -ray source, both calibrated with ion chambers. The dosemeters have been used as an independent check of the output calibration of a linear accelerator before clinical use. The absorbed dose of the dosemeters was determined by comparison with those exposed to a 60 Co gamma -ray source calibrated with Fricke dosimetry.