Key result
The proposed short-term post-event ECG recorder provided clinically acceptable signals for diagnosis in 96% of cases, demonstrating statistical equivalence to the 98% acceptability of the golden standard 12-lead ECG.
Why the study?
Does a novel patient-activated short-term post-event ECG recorder provide acceptable signal quality for diagnosis compared to standard 12-lead ECG in patients?
Observational (n=391)
Single-blind
No
Does a novel patient-activated short-term post-event ECG recorder provide acceptable signal quality for diagnosis compared to standard 12-lead ECG in patients?
A novel multi-purpose ECG telemetry system provides reliable short-term post-event recordings that are acceptable for diagnosis in 96% of cases.
Answers are grounded in this paper and cite their sources.
Supports feasibility of patient-activated telemetry for symptomatic arrhythmia monitoring; leaves open clinical utility and diagnostic yield in.
BACKGROUND: The Electrocardiogram ECG is one of the most important non-invasive tools for cardiac diseases diagnosis. Taking advantage of the developed telecommunication infrastructure, several approaches that address the development of telemetry cardiac devices were introduced recently. Telemetry ECG devices allow easy and fast ECG monitoring of patients with suspected cardiac issues. Choosing the right device with the desired working mode, signal quality, and the device cost are still the main obstacles to massive usage of these devices. METHODS: In this paper, we introduce design, implementation, and validation of a multi-purpose telemetry system for recording, transmission, and interpretation of ECG signals in different recording modes. The system consists of an ECG device, a cloud-based analysis pipeline, and accompanied mobile applications for physicians and patients. The proposed ECG device's mechanical design allows laypersons to easily record post-event short-term ECG signals, using dry electrodes without any preparation. Moreover, patients can use the device to record long-term signals in loop and holter modes, using wet electrodes. In order to overcome the problem of signal quality fluctuation due to using different electrodes types and different placements on subject's chest, customized ECG signal processing and interpretation pipeline is presented for each working mode. RESULTS: We present the evaluation of the novel short-term recorder design. Recording of an ECG signal was performed for 391 patients using a standard 12-leads golden standard ECG and the proposed patient-activated short-term post-event recorder. In the validation phase, a sample of validation signals followed peer review process wherein two experts annotated the signals in terms of signal acceptability for diagnosis.We found that 96% of signals allow detecting arrhythmia and other signal's abnormal changes. Additionally, we compared and presented the correlation coefficient and the automatic QRS delineation results of both short-term post-event recorder and 12-leads golden standard ECG recorder. CONCLUSIONS: The proposed multi-purpose ECG device allows physicians to choose the working mode of the same device according to the patient status. The proposed device was designed to allow patients to manage the technical requirements of both working modes. Post-event short-term ECG recording using the proposed design provide physicians reliable three ECG leads with direct symptom-rhythm correlation.
No takes yet. Share an insight, caveat, or question.
Marouf et al. (2017) conducted an observational in Non-significant cardiac issues (n=391). Multi-purpose ECG telemetry system (short-term post-event recorder) vs. Standard 12-lead golden standard ECG was evaluated on Clinical acceptability of signals for diagnosis (p=0.796). The proposed short-term post-event ECG recorder provided clinically acceptable signals for diagnosis in 96% of cases, demonstrating statistical equivalence to the 98% acceptability of the golden standard 12-lead ECG.
Synapse has enriched 10 related papers. Consider them for comparative context: