Key result
Atrial fibrillation did not significantly affect the accuracy of oscillometric blood pressure measurements compared to sinus rhythm when 3 repeated measurements were performed (P>0.05).
Why the study?
Does atrial fibrillation affect the accuracy of automated oscillometric blood pressure monitoring compared to invasive arterial pressure?
Cross-Sectional (n=102)
Does atrial fibrillation affect the accuracy of automated oscillometric blood pressure monitoring compared to invasive arterial pressure?
p-value: p=>0.05
Atrial fibrillation does not significantly affect the accuracy of oscillometric blood pressure measurements when the mean of 3 repeated measurements is used.
May support averaged oscillometric BP in AF; leaves open prospective validation across devices and populations.
The introduction of automated oscillometric blood pressure monitors was the basis for today's widespread use of blood pressure self-measurement. However, in atrial fibrillation, there is a controversial debate on the use of oscillometry because there is a high variability of heart rate and stroke volume. To date, the accuracy of oscillometric blood pressure monitoring in atrial fibrillation has only been investigated using auscultatory sphygmomanometry as reference method, which may be biased by arrhythmia as well. We performed a cross-sectional study in 102 patients (52 sinus rhythm, 50 atrial fibrillation) assessing the accuracy of an automated and validated oscillometric upper arm (M5 Professional, Omron) and wrist device (R5 Professional, Omron) to invasively assessed arterial pressure. Blood pressure values were calculated as the mean of 3 consecutive measurements. Systolic and diastolic blood pressure did not significantly differ in patients with sinus rhythm and atrial fibrillation, independent of the method of measurement (P>0.05 each). The within-subject variability of the oscillometric measurements was higher in patients with atrial fibrillation compared with sinus rhythm (P<0.01 each). The biases of systolic and diastolic blood pressure, however, did not significantly differ in presence or absence of atrial fibrillation in Bland-Altmann analysis (P>0.05 each). In conclusion, atrial fibrillation did not significantly affect the accuracy of oscillometric measurements, if 3 repeated measurements were performed.
No takes yet. Share an insight, caveat, or question.
Pagonas et al. (2013) conducted a cross-sectional in Atrial fibrillation (n=102). Atrial fibrillation vs. Sinus rhythm was evaluated on Bias of systolic and diastolic blood pressure compared to invasive arterial pressure (p=>0.05). Atrial fibrillation did not significantly affect the accuracy of oscillometric blood pressure measurements compared to sinus rhythm when 3 repeated measurements were performed (P>0.05).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: