Key result
Atrial fibrillation impacts the accuracy of automated oscillometric blood pressure measurements, with diastolic bias exceeding 5 mm Hg.
This response clarifies that the underestimation of systolic and overestimation of diastolic blood pressure is a general limitation of oscillometry, independent of atrial fibrillation, and emphasizes the need to include elderly subjects in validation protocols.
Response to Does Atrial Fibrillation Affect the Automated Oscillometric Blood Pressure Measurement?We thank Stergiou and coworkers for their interest in our study.1 The authors criticize that we did not use a validation protocol.The present study examines the impact of one single parameter (atrial fibrillation, AF) on the accuracy of oscillometry.It is not a validation study.The performance of a device in a validation procedure is determined by a broad variety of parameters.Bland-Altman analysis-as used in our study-is the statistical gold standard to assess the accuracy of a measurement technique and allows to analyze the impact of a single parameter like AF by statistical comparison of the resulting biases. 2 Furthermore, the statement "the AAMI (Association for Advancement of Medical Instrumentation) 3 criterion of accuracy was fulfilled in SR but not in AF" is only true for systolic blood pressure, because the diastolic bias exceeded 5 mm Hg.Concerning Stergiou's question of a potentially underpowered study regarding systolic pressure in AF, it may be of interest that the number of measurements per patient was in the range of the requested number of measurements in the AAMI protocol (150 measurements).The following 2 issues have to be clearly differentiated: first, the impact of AF on the accuracy of measurements; second, the problem that blood pressure monitors that have successfully passed validation protocols fail to prove the same accuracy in clinical practice-as was evident in our study.It has to be kept in mind that prior validations of the devices used sphygmomanometry as reference.Moreover, validation protocols unfortunately do not mandatorily require the inclusion of elderly subjects with a higher degree of arteriosclerosis.We have previously shown that the accuracy of oscillometry is impaired by an increase of pulse pressure as a footprint of vascular ageing.4 As mentioned in that article, we, therefore, strongly recommend the mandatory inclusion of elderly subjects in future validation protocols.The authors remarked the high bias of pulse pressure.The underestimation of systolic and overestimation of diastolic blood pressure is a general problem of oscillometry that has been repeatedly described independent of AF. 5 We agree with the authors that the results of 2 single devices cannot be extrapolated to the oscillometric technique in general.Therefore, we want to repeat the last sentence of the discussion: "The study should be repeated using oscillometric devices from Letter to the Editor (Hypertension.
No takes yet. Share an insight, caveat, or question.
Pagonas et al. (2013) conducted a letter in Atrial fibrillation. Atrial fibrillation vs. Sinus rhythm was evaluated on Accuracy of oscillometric blood pressure measurement. Atrial fibrillation impacts the accuracy of automated oscillometric blood pressure measurements, with diastolic bias exceeding 5 mm Hg.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: