Key result
An amperometric immunosensor developed using a gold CD-R transducer presented a cut-off value of -0.949 µA for the determination of Chagas' disease in sera samples.
A novel amperometric immunosensor using a gold CD-R transducer was developed for the detection of Chagas' disease in patient sera.
Answers are grounded in this paper and cite their sources.
May support low-cost Chagas serology in limited settings; leaves open prospective clinical validation.
Chagas’ disease has been spread mainly through blood transfusion or blood components, therefore the development of a device for neglected diseases with high specificity and sensitivity is important to public health. This work describes the development of an amperometric immunosensor for determination of Chagas’ disease through a gold‐based electrode obtained from a recordable compact disc modified with 4‐(methylmercapto)benzaldehyde for the immobilization of Tc85 protein of the Trypanosoma cruzi . The immunoassays were carried out using positive and negative sera from Chagas’disease patients and immunoglobulin conjugated with peroxidase enzyme. The immunosensor presented −0.949 µA as cut‐off value and was applied in sera samples.
No takes yet. Share an insight, caveat, or question.
Foguel et al. (2011) studied Chagas' disease. Amperometric immunosensor using gold CD-R transducer vs. Positive and negative sera was evaluated on Determination of Chagas' disease (cut-off value). An amperometric immunosensor developed using a gold CD-R transducer presented a cut-off value of -0.949 µA for the determination of Chagas' disease in sera samples.