The CERES instrument has three radiometers which operate in the short-wave (0,3 μm to 5 μm), the long-wave (8 μm to 12 μm) and total broadband (0,3 μm to >200 μm) spectral regions. The long-wave calibration plan for CERES provides for absolute radiometric calibration of the two CERES sensors operating in the long-wave portion of the spectrum. Ground calibration is carried out in the TRW Radiometric Calibration Facility (RCF), a cryogenic vacuum chamber with the capability to simulate radiometrically both cold space and Earth viewing conditions. Absolute calibration is made using a temperature-based radiometric scale created by a black-body radiance standard controlled at temperatures from 200 K to 320 K. In May 1992, the CERES radiometric test model (RTM), a developmental but functionally flight-like instrument, was calibrated at TRW in the RCF. This was the first end-to-end long-wave calibration of a CERES sensor and provided a test and initial validation of the long-wave calibration plan and procedures. The gain, offset, out-of-field response and spectral response were determined for each sensor. The end-to-end long-wave calibration plan is described and the RTM calibration and validation results are presented.