Answers are grounded in this paper and cite their sources.
The response of radiochromic film, GafChromic EBT, was investigated for dependence on x-ray beam energy using a previously reported real-time optical readout approach. X-ray beams of energy from 75 kVp to 18 MV were employed. The dose-induced change in optical density for the EBT film was compared to values obtained for GafChromic HS and MD-55 films, exposed under the same conditions. All responses were normalized to that obtained for 60Co irradiation. While change in optical density for 1 Gy of applied dose as measured with HS and MD-55 films decreased by approximately 40% at low energies, the mean change in optical density of EBT film remained within 3% of that in the 60Co beam over the entire energy range.
No takes yet. Share an insight, caveat, or question.
Rink et al. (2007) studied this question.