A sudden decrease in right ventricular electrogram amplitude in an ICD lead can be caused by a new right bundle branch block altering the activation wavefront, mimicking lead malfunction.
A 76-year-old male had dual-chamber implantable cardioverter defibrillatorimplanted, employing a Medtronic Sprint Fidelis 6931-65 shock. A routine follow-up 4 years later showed a sudden steep decrease in the amplitude of the right ventricular electrogram (EGM) from 9.0 to 2.0 mV (Panel A). Because of the known high failure rate related to the Sprint Fidelis shock lead, we initially focused on markers of lead failure by recordings of the intracardiac EGM, changes in lead impedance, and/or radiographic abnormalities. However, in none of these parameters any abnormality was observed. Lead impedance trend was completely stable at the time the decrease in EGM amplitude was observed. R-wave amplitude showing a decrease of the R wave from 9.0 to 2.0 mV (A). Simultaneous recording of the lead impedance trend shows no change in the same period (B). Electrocardiogram when R-wave amplitude was 9.0 mV (C). The electrocardiogram 6 months later shows appearance of a right bundle branch block after a decrease of the R-wave amplitude (D). Because no explanation was found from these parameters and lead malfunction was excluded, we compared the recent surface electrocardiogram (ECG) with previous recordings (Panel B). In the latter ECG the QRS had a normal configuration, whereas the recent recording showed a right bundle branch block (Panel C), which caused the decreased amplitude. The difference in activation wavefront resulting from altered conduction at the site of block therefore influences the direction of the local vector of depolarization. The acquired EGM is the projection of this vector on the dipole formed by the sensing electrodes of the shock lead, and the amplitude may be greatly diminished when the activation wavefront arrives more perpendicular to the dipole. In case of a sudden decrease in amplitude of the EGM, a change in the intrinsic conduction should be considered as the possible cause. To the best of our knowledge this observation has not been reported previously. Conflict of interest: B.M.v.G. served as Clinical Advisor, Medtronic Trading, NL BV, Heerlen, the Netherlands, and as a Consultant in RADI Pressure Wire Systems, a Sint Jude Medical company. R.N. and F.A.B. have no conflict of interest. The full-length version of this report can be viewed at: http://www.escardio.org/communities/EHRA/publications/ep-case-reports/Documents/electrogram-amplitude-decrease.pdf
No takes yet. Share an insight, caveat, or question.
Gelder et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: