What is the incidence and what are the predictors of permanent pacemaker requirement following TAVR with the Lotus valve in patients with severe aortic stenosis?
What is the incidence and what are the predictors of permanent pacemaker requirement following TAVR with the Lotus valve in patients with severe aortic stenosis?
Following TAVR with the Lotus valve, 28% of patients required a permanent pacemaker within 30 days, with pre-existing RBBB and deeper valve implantation being significant predictors.
High pacemaker rates after Lotus TAVR warrant pre-procedural risk stratification; leaves open whether optimized implantation reduces need in prospective trials.
OBJECTIVES: Valve System (Boston Scientific). BACKGROUND: Pacemaker implantation is the most common complication following TAVR. Predictors of pacing following TAVR with the Lotus valve have not been systematically assessed. METHODS: Consecutive patients with severe aortic stenosis who underwent Lotus valve implantation were prospectively recruited at a single-centre. Patients with a pre-existing PPM were excluded. Baseline ECG, echocardiographic and multiple detector computed tomography as well as procedural telemetry and depth of implantation were independently analyzed in a blinded manner. The primary endpoint was 30-day incidence of pacemaker requirement (PPM implantation or death while pacing-dependent). Multivariate analysis was performed to identify independent predictors of the primary endpoint. RESULTS: A total of 104 consecutive patients underwent TAVR with the Lotus valve with 9/104 (9%) with a pre-existing PPM excluded. New or worsened procedural LBBB occurred in 78%. Thirty-day incidence of the primary pacing endpoint was 28%. The most common indication for PPM implantation was complete heart block (CHB) (69%). Independent predictors of the primary endpoint included pre-existing RBBB (hazard ratio [HR] 2.8, 95% CI 1.1-7.0; P = 0.032) and depth of implantation below the noncoronary cusp (NCC) (HR 2.4, 95% CI 1.0-5.7; P = 0.045). CONCLUSIONS: Almost a third of Lotus valve recipients require pacemaker implantation within 30 days. The presence of pre-existing RBBB and the depth of prosthesis implantation below the NCC were significant pacing predictors.
No takes yet. Share an insight, caveat, or question.
Zaman et al. (2016) studied this question.